QChKAdC
QChKAdC is a desktop diamond verification instrument for loose and mounted diamonds, designed to separate diamonds from synthetic diamonds. QChKAdC is manually operated and requires an operator to interpret the test results.
QChKAdC is a desktop diamond verification instrument for loose and mounted diamonds, designed to separate diamonds from synthetic diamonds. QChKAdC is manually operated and requires an operator to interpret the test results.
Weight: 0.003 ct– 0.2 ct
Size: 0.86 mm-3.7 mm
Colour: D-J
Shape: All
Diamond Simulants: No
Mounted Stones: Yes (not yet ASSURE Tested)
The below results demonstrate the performance of the QChKAdC with the ASSURE Core Sample:
Diamond False Positive Rate (optimal rate 0%): 11.1%
Diamond Referral Rate (optimal rate 0%): N/A
Diamond Accuracy Rate (optimal rate 100%): 94.2%
Speed (stones tested per hour): 731
The ASSURE Core Sample consists of 1,000 loose diamonds, 200 loose synthetic diamonds and 200 loose diamond simulants. The stones are 0.02ct to 0.20ct in D-J colour.
The below results demonstrate the performance of the QChKAdC with the ASSURE Smalls Sample:
Diamond False Positive Rate (optimal rate 0%): 3.7%
Diamond Referral Rate (optimal rate 0%): N/A
Diamond Accuracy Rate (optimal rate 100%): 88.6%
Speed (stones tested per hour): 731
The ASSURE Smalls Sample consists of 1,000 loose diamonds, 200 loose synthetic diamonds and 200 loose diamond simulants. The stones are 0.005 ct to 0.02 ct in D-J colour.
Automated Feed: No
Automated Results: No
Automated Dispense: No
Detect or Refer (Synthetic Diamonds): Detect
Detect or Refer (Diamond Simulants): N/A
Multiple Stones at Once: Yes
Training Required: Yes
After loading the sample in the aluminium plate, it is exposed to deep UV and its phosphorescence and fluorescence are monitored in the windows. The Phosphorescence of some of the synthetic diamonds can be enhanced when the sample is kept at slightly higher temperature (approx. 60-70 degrees C) which can be easily executed by placing the sample loaded metal plate on a vessel containing hot water.
QChKAdC makes a measurement and displays one of the following results:
From the observations and inference regarding Natural/Synthetic is by manual interpretation of the colour of Fluorescence and phosphorescence as per the instruction manual supplied.
Size: 35.5 cm (W) x 43 cm (D) x 49 cm (H)
Weight: 14.5 kg
Price: 7,700 USD excl. laptop (September 2019)
quickcheck.co.in/
enquiry@arotek-inst.com
+9122 25301119
ASSURE Tested September 2019
ASSURE ID: 48396